FUNCTIONAL:
In accordance with MIL-P-24212C (SH)
APPLICATIONS: Steam, oil, sea water, flue
gas, ammonia, oxygen, other gases, fresh
water, condensate
OUTPUT: 4-20 mADC
POWER SUPPLY: 28 VDC with reverse protection.
LOOP RESISTANCE: 0-250 ohms
ZERO/SPAN ADJUSTMENTS: 8% F.S. output
TEMPERATURE LIMITS: Compensated: 0°C to
65°C (32°F to 149°F)
OVER PRESSURE: 200% of range up to maximum of 10,000 psi
(whichever is less)
HUMIDITY: 0-100% RH
VOLUMETRIC DISPLACEMENT: less than 0.002 cc
ACCURACY: ± 1.0% of calibrated span including
linearity, hysteresis, and temperature effects
REPEATABILITY: within 0.5% of span
SHOCK TEST: in accordance with Grade-A, Class I, Type C of
MIL-S-901
EMI: in accordance with MIL-P-24212C
(SH) |